Medtronic Perfusion Systems — Octopus Evolution AS Tissue Stabilizer, Model TS2500

During the manufacturing process, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.

M
Medtronic Perfusion Systems — Octopus 4 Tissue Stabilizer, Model 29400

During the manufacturing process, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.

M
Medtronic Perfusion Systems — Octopus Evolution Tissue Stabilizer, Model TS2000

During the manufacturing process, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.

F
Fresenius Medical Care Holdings, Inc. — Bicarby" Dialysate; Model Numbers: RFP-400-G, RFP-401-G, RFP-402-G, RFP-404-G, RFP-407-G, and RFP-456-G;

Given the increased number of leaks reported by users, it was determined to be the appropriate course of action to notify users of the potential for the occurrence of Bicarby luer-lock leaks during manipulation of the frangible and tubing when preparing the product for use for treatment. This issue presents the potential of a slip and fall hazard associated with fluid on the floor.

B
Boston Scientific Corporation — Orca Single Use Air/Water and Suction Valves, Material Number (UPN) SUV-617-50; The Orca Air/Water Valve is intended to be used to control the air/water function on an

Boston Scientific is initiating a medical device removal of Orca Sterile, Single Use Air/Water and Suction Valves due to a higher than anticipated occurrence of the suction button sticking.

A
Abbott Medical — Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software

Software malfunction may occur during pacing capture threshold test of pacemaker resulting in programmer not successfully communicating the command to terminate the test.

H
Haleon US Holdings LLC — Gas-X, Simethicone 125 mg/ ANTIGAS, 120 SoftGels, Distributed by: Haleon, Warren, NJ 07059, UPC: 3 00674 35041 9.

Labeling: Not Elsewhere Classified- Incomplete inactive ingredient information on the carton labeling.

I
Inspire Medical Systems Inc. — Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire stimulation Lead Model/Catalog Number: 4063 Product Description: The Inspire¿ Stimulation Lead (Model

Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a product past its Use By Date may result in unplanned or premature revision surgery.

B
Boston Scientific Corporation — Intera Oncology INTERA 3000 Hepatic Artery Infusion Pump, AP-0300H, implantable drug delivery device

There is the potential for leakage along the drug pathway from the pump through the end of the catheter.

Z
I
Inspire Medical Systems Inc. — Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire Respiratory Sensing Lead Model/Catalog Number: 4340 Software Version: N/A Product Description: The In

Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a product past its Use By Date may result in unplanned or premature revision surgery.

J
J and J Bunting, Inc — Lunds & Byerly's Curry Powder Sweet

Sesame Seeds White were labeled as Curry Powder Sweet.

V
Ventec Life Systems, Inc. — VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001

Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition.

D
Datex-Ohmeda Inc — Giraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator) Product identifiers: 2082663-001, 2082663-002, M1241424 Product impacted only if ser

Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.

D
Datex-Ohmeda Inc — Giraffe Warmer with Integrated Resuscitation System (iRes), Panda Warmer with Integrated Resuscitation System (iRes) Product Identifiers: M1118179, M1112198 Impacted only if servic

Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.

G
GE Medical Systems SCS — GE Healthcare AW Server 3.2 ext.6.5; Radiological Image Processing System

GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches an interactive application (e.g. Volume Viewer), the application may open the previous patient's exam instead of the intended one. When this issue occurs, there is no system warning or error notification. If the error is not recognized, a clinical user could review, interpret, or report images for the wrong patient which could lead to misdiagnosis, incorrect clinical decisions, resulting in delayed or incorrect treatment.

D
Datex-Ohmeda Inc — Panda Freestanding Warmer, Infant Radiant Warmer Product Identifiers: 2063568-001 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance

Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.

A
AMD Medicom Inc. — Ritmed Neurological Sponges, Description/REF: 0.5x0.5in, Non-Sterile/55992C; 0.5x1.0in, Non-Sterile/55993C; 0.5x3.0in, Non-Sterile/55994C; 0.75x0.75in, Non-Sterile/55995C;

Neurological sponges used to moisten or absorb biological fluid during neurological surgery and may also be used to protect the brain during incision, may not consistently meet endotoxin specification for devices that contact the cerebrospinal fluid. Elevated endotoxin levels may lead to neurovascular adverse events, febrile response, and/or local transient inflammation, hypotension, or nausea.

A
AMD Medicom Inc. — Ritmed Neurological Sponges, 0.75x0.75in, Sterile, REF: 85995C

Neurological sponges used to moisten or absorb biological fluid during neurological surgery and may also be used to protect the brain during incision, may not consistently meet endotoxin specification for devices that contact the cerebrospinal fluid. Elevated endotoxin levels may lead to neurovascular adverse events, febrile response, and/or local transient inflammation, hypotension, or nausea.

F
Fresenius Medical Care Holdings, Inc. — Bicarby Dialysate; Model number: RFP-400-G;

There is an error in the labeling of the GTIN on the product. When scanned; however, the case labeling, lot number, and printed labeling on the product correctly identifies the product code and the solution contained in the bag. In the event the incorrect GTIN labeling is used as the singular source for product identification, unrecognized use of a potassium solution with a concentration differing from the prescription may result.

D
Draeger, Inc. — Atlan A350XL. Model Number: 8621600. anesthesia workstation

The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.

S
Siemens Healthcare Diagnostics, Inc. — Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay

A potential positive bias in quality control (QC) and patient sample results in three lots of the Dimension Tacrolimus (TAC) Flex reagent cartridge.

D
Draeger, Inc. — Atlan A350. Model Number: 8211500. anesthesia workstation

The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.