GE Medical Systems SCS — GE Healthcare AW Server 3.2 ext.6.5; Radiological Image Processing System

⚠️ GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches an interactive application (e.g. Volume Viewer), the application may open the previous patient's exam instead of the intended one. When this issue occurs, there is no system warning or error notification. If the error is not recognized, a clinical user could review, interpret, or report images for the wrong patient which could lead to misdiagnosis, incorrect clinical decisions, resulting in delayed or incorrect treatment.

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
May 8, 2026

Worldwide distribution - United States Nationwide and the countries of Algeria, Australia, Belarus, Belgium, Benin, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Korea (Republic Of), Lithuania, Morocco, Nepal, Netherlands, New Zealand, North Macedonia, Norway, Pakistan, Poland, Portugal, Romania, Russia, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkiye, United Arab Emirates, United Kingdom, Vietnam.

What's affected?

Product
GE Healthcare AW Server 3.2 ext.6.5; Radiological Image Processing System
Lot code
REF# (01)00840682102384(10)AWS03D02E6D5, GTIN Number 00840682102384

Additional details

REF# (01)00840682102384(10)AWS03D02E6D5, GTIN Number 00840682102384

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 340 servers.