Recall
FDAHealthMay 8, 2026

AMD Medicom Inc. — Ritmed Neurological Sponges, Description/REF: 0.5x0.5in, Non-Sterile/55992C; 0.5x1.0in, Non-Sterile/55993C; 0.5x3.0in, Non-Sterile/55994C; 0.75x0.75in, Non-Sterile/55995C;

⚠️ Neurological sponges used to moisten or absorb biological fluid during neurological surgery and may also be used to protect the brain during incision, may not consistently meet endotoxin specification for devices that contact the cerebrospinal fluid. Elevated endotoxin levels may lead to neurovascular adverse events, febrile response, and/or local transient inflammation, hypotension, or nausea.

Is this relevant to me?

Geography
US, MT, NC, VA, CA, TX, OH, TN, IL, RI
Source agency
FDA
Category
Health
Published
May 8, 2026

US: MT, NC, VA, CA, TX, OH, TN, IL, RI

What's affected?

Product
Ritmed Neurological Sponges, Description/REF: 0.5x0.5in, Non-Sterile/55992C; 0.5x1.0in, Non-Sterile/55993C; 0.5x3.0in, Non-Sterile/55994C; 0.75x0.75in, Non-Sterile/55995C; 1.0x1.0in, Non-Sterile/55996C; 1.0x3.0in, Non-Sterile/55997C; 0.5x1.5in, Non-Sterile/55998C
Lot code
19316

Additional details

UDI-DI: 20686864043205. Distributed After May 1st, 2022, including the following lots: 19316, 16409, 15463-7, 15463-6, 15463-5, 15463-4, 15463-3, 15463-2, 15463-1, 15463, 15192-1, 15192, 14813, 14039, 13673, 13289, 12641, 12206, 11851, 11020-1, 11020, 10586, 10392, 9243, 8160, 7784, 7748, 7044

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 7,838,200.

View original FDA notice →