Medline Industries, LP — MEDLINE Medical Procedure Kits labeled as: CATH 22F ROBINSON STRL 5'S PK, Medline Kit Number/SKU DYNJ0040839
⚠️ During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Is this relevant to me?
- Geography
- Nationwide
- Source agency
- FDA
- Category
- Health
- Published
- May 4, 2026
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
What's affected?
- Product
- MEDLINE Medical Procedure Kits labeled as: CATH 22F ROBINSON STRL 5'S PK, Medline Kit Number/SKU DYNJ0040839
- Lot code
- 25JMK709
Additional details
UDI/DI 10889942046126 (each), 40889942046127 (case), Lot Numbers: 25JMK709, 25FMJ661, 25CMG042, 24IME549, 24GMD004, 24CMF091, 24CME465, 24BMH490.
What should I do?
Recommended action
Class II recall. Voluntary: Firm initiated Quantity affected: 1000 units.
Get Health recall alerts
Delivered to your inbox. Free. No account required.