medline industries, lp
50 recalls on record
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
Expansion of RES 98277. Specific lots of reprocessed catheters may contain small particles of residual material. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection. If particulate were to dislodge during use and contact circulating blood or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.
The Medline Convenience Kits contain Arrow Kits and Sets that include Lidocaine Hydrochloride Injection, USP and Bupivacaine Hydrochloride in Dextrose Injection, USP which were recalled due to quality issues identified during a recent FDA inspection of the supplier.
The Medline Convenience Kits contain Arrow Kits and Sets that include Lidocaine Hydrochloride Injection, USP and Bupivacaine Hydrochloride in Dextrose Injection, USP which were recalled due to quality issues identified during a recent FDA inspection of the supplier.
The Medline Convenience Kits contain Arrow Kits and Sets that include Lidocaine Hydrochloride Injection, USP and Bupivacaine Hydrochloride in Dextrose Injection, USP which were recalled due to quality issues identified during a recent FDA inspection of the supplier.
The Medline Convenience Kits contain Arrow Kits and Sets that include Lidocaine Hydrochloride Injection, USP and Bupivacaine Hydrochloride in Dextrose Injection, USP which were recalled due to quality issues identified during a recent FDA inspection of the supplier.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.