Recall
FDAHealthApril 27, 2026

Medline Industries, LP — Medline convenience kit: BHS NEURO DYNJ904900N MEDTRX NEURO PACK DYNJ67267B MRI DBS PACK DYNJ9814630O NEURO CDS-LF CDS983383L NEURO PACK DYNJ59397Q DYNJT7

⚠️ Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
April 27, 2026

US Nationwide distribution. OUS distribution pending.

What's affected?

Product
Medline convenience kit: BHS NEURO DYNJ904900N MEDTRX NEURO PACK DYNJ67267B MRI DBS PACK DYNJ9814630O NEURO CDS-LF CDS983383L NEURO PACK DYNJ59397Q DYNJT7315 NEURO PACK-LF DYNJ25097I NEURO VP SHUNT/VAGAL NERVE-RF DYNJ21925S PK NEURO SHUNT BUMC DYNJT6428 RR-NEURO DBS HEAD PACK-LF DYNJ0785736M VP SHUNT SYNJ10307B
Lot code
25LBB457

Additional details

DYNJ904900N UDI-DI 10198459504273 Lot 25LBB457 DYNJ67267B UDI-DI 10198459084331 Lot 25LBQ732 DYNJ9814630O UDI-DI 10195327587291 Lot 25LBD665 CDS983383L UDI-DI 10198459224577 Lot 25LBC092 DYNJ59397Q UDI-DI 10198459495427 Lot 25LMG049 DYNJT7315 UDI-DI 10198459620225 Lot 25LBO095 DYNJ25097I UDI-DI 10198459449550 Lot 25KBC747 DYNJ21925S UDI-DI 10195327105020 Lots 25KBD835 25LBR371 DYNJT6428 UDI-DI 10198459553660 Lot 26AMA602 DYNJ0785736M UDI-DI 10198459487149 Lot 25LMI831 SYNJ10307B UDI-DI 10195327007249 Lot 25LBE309

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 49,654 kits total.

View original FDA notice →

Get Health recall alerts

Delivered to your inbox. Free. No account required.