Medline Industries, LP — Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: TAVR PACK, Medline SKU # DYN
⚠️ Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Is this relevant to me?
- Geography
- Nationwide
- Source agency
- FDA
- Category
- Health
- Published
- February 27, 2026
Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).
What's affected?
- Product
- Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: TAVR PACK, Medline SKU # DYNJ84472
- Lot code
- 23LBN016
Additional details
Medline SKU # DYNJ84472, UDI/DI each 10195327369286, UDI/DI case 40195327369287, Lot Numbers: 23LBN016, 23KBL255, 23HBL885, 23FBV799.
What should I do?
Recommended action
Class I recall. Voluntary: Firm initiated Quantity affected: 108 units.