Medline Industries, LP — THORACIC ROBOTS, DYNJ908777B
⚠️ Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
Is this relevant to me?
- Geography
- Nationwide
- Source agency
- FDA
- Category
- Health
- Published
- February 16, 2026
US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.
What's affected?
- Product
- THORACIC ROBOTS, DYNJ908777B
- Lot code
- 25EBD732
Additional details
UDI-DI 10198459269196 (EA) 40198459269197 (CS) LOTS 25EBD732 25GBG733 25IBQ613 25JBD457 25KBT352 26ABA235 26ABP735
What should I do?
Recommended action
Class II recall. Voluntary: Firm initiated Quantity affected: 14,379 kits total.
Get Health recall alerts
Delivered to your inbox. Free. No account required.