Recall
FDAHealthJanuary 8, 2026

Medline Industries, LP — Convenience kits containing Tego Connectors Medline DIALYSIS DRSNG CHNG SKU DYNDC3167

⚠️ Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
January 8, 2026

US Nationwide distribution.

What's affected?

Product
Convenience kits containing Tego Connectors Medline DIALYSIS DRSNG CHNG SKU DYNDC3167
Lot code
24IME953

Additional details

UDI-DI (ea) 10195327015954 UDI-DI (case) 40195327015955 Kit lots 24IME953 25CMA708

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 150.

View original FDA notice →

Get Health recall alerts