Recall
FDAHealthJanuary 8, 2026

Medline Industries, LP — Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746

⚠️ Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
January 8, 2026

US Nationwide distribution.

What's affected?

Product
Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746
Lot code
2024051380

Additional details

UDI-DI (ea) 10653160373582 UDI-DI (case) 00653160373585 Kit lots 2024051380 2024052080 2024052180 2024052280 2024071580 2025061790 2025072990

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 2020.

View original FDA notice →

Get Health recall alerts

Delivered to your inbox. Free. No account required.