Medline Industries, LP — Intubation Tray (Adult) containing Suction Catheter Kits SKU DYNJAA245B
⚠️ A smaller open suction catheter, the 8 Fr Suction Catheter, may have been packed within the kit instead of the required 14 Fr Suction Catheter. Use of the smaller size catheter may result in hypoxia and/or respiratory failure related to partial or full airway obstruction. Airway obstruction may require immediate emergent medical intervention to prevent respiratory failure or death. Other potential outcomes include prolonged procedure, airway irritation, or prolonged dyspnea.
Is this relevant to me?
- Geography
- Nationwide
- Source agency
- FDA
- Category
- Health
- Published
- December 24, 2025
Distribution US nationwide and Canada.
What's affected?
- Product
- Intubation Tray (Adult) containing Suction Catheter Kits SKU DYNJAA245B
- Lot code
- 25IBC172
Additional details
Kit lots 25IBC172 25HBJ413 UDI (ea) 10198459480317 UDI (case) 40198459480318
What should I do?
Recommended action
Class II recall. Voluntary: Firm initiated Quantity affected: 130.