Medline Industries, LP — MEDLINE beds labeled as: Full Electric Lightweight Homecare Bed, REF MDR107003L
⚠️ Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.
Is this relevant to me?
- Geography
- US, CA, PA, VI
- Source agency
- FDA
- Category
- Health
- Published
- November 26, 2025
US, CA, PA, GUAM, VI
What's affected?
- Product
- MEDLINE beds labeled as: Full Electric Lightweight Homecare Bed, REF MDR107003L
- Lot code
- s
Additional details
UDI/DI 40080196294344, All lots
What should I do?
Recommended action
Class I recall. Voluntary: Firm initiated Quantity affected: 35694 units.