Recall
FDAHealthNovember 26, 2025

Medline Industries, LP — MEDLINE beds labeled as: Semi Electric Lightweight Homecare Bed, REF MDR107002L

⚠️ Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.

Is this relevant to me?

Geography
US, CA, PA, VI
Source agency
FDA
Category
Health
Published
November 26, 2025

US, CA, PA, GUAM, VI

What's affected?

Product
MEDLINE beds labeled as: Semi Electric Lightweight Homecare Bed, REF MDR107002L
Lot code
s

Additional details

UDI/DI 40080196294320, All lots

What should I do?

Recommended action

Class I recall. Voluntary: Firm initiated Quantity affected: 13245 units.

View original FDA notice →

Get Health recall alerts

Delivered to your inbox. Free. No account required.