Vision RT Ltd — The AlignRT InBore system contains six Class 1 lasers which enables the system s cameras to track the patient s position and movement.

Vision RT is informing customers of an omission identified in the Instructions for Use for AlignRT InBore systems. Information identifying the lasers and their characteristics was omitted from the Instructions for Use for the subject system as required by 21 CFR 1040,10(h), via conformance with recognized consensus standard IEC 60825-1 Ed. 3 (FDA Laser Notice 56).

B
B BRAUN MEDICAL INC — STERILE WATER FOR INJECTION USP, 3000mL, Rx only, B. Braun Medical Inc., Bethlehem. PA 18018, NDC 0264-7385-60.

Lack of Assurance of Sterility- Potential for fluid leakage from the port due to port misalignment.

W
W L Gore & Associates, Inc. — GORE ACUSEAL Vascular Graft, REF: ECH050020J, ECH050020W, ECH050050J, ECH050050W, ECH060010A, ECH060020A, ECH060020J, ECH060020W, ECH060040, ECH060040A, ECH060040W, ECH06

Reports of vascular graft delamination, which may lead to reduced access flow; cannulation difficulties; bleeding or bruising; and harms related to reintervention procedures, so Instructions for Use will be updated to modify existing warning to 1) further clarify techniques that may contribute to the risk of delamination, 2) Addition of delamination to device-related adverse event section.

O
OSSTEM Implant Co., Ltd. — Osstem Dental X-Ray System T2. Product Model Numbers T2-C and T2-CS.

Osstem Implant Co. Ltd, is voluntarily initiating a medical device correction involving the current labeling for this electronic product may be missing a required Certification Statement under 21 CFR Subchapter J (Radiological Health Regulations). Specifically, the following statement may have been omitted: Certification Statement: This electronic product complies with all applicable performance standards prescribed under 21 CFR Subchapter J, as enforced by the U.S. Food and Drug Administration at the date of manufacturing.

O
OSSTEM Implant Co., Ltd. — Osstem Dental X-Ray System T1. Product Model Numbers T1-C and T1-CS.

Osstem Implant Co. Ltd, is voluntarily initiating a medical device correction involving the current labeling for this electronic product may be missing a required Certification Statement under 21 CFR Subchapter J (Radiological Health Regulations). Specifically, the following statement may have been omitted: Certification Statement: This electronic product complies with all applicable performance standards prescribed under 21 CFR Subchapter J, as enforced by the U.S. Food and Drug Administration at the date of manufacturing.

O
OSSTEM Implant Co., Ltd. — Osstem Dental X-Ray System T2Plus. Product Model Numbers T2-C-P and T2-CS-P.

Osstem Implant Co. Ltd, is voluntarily initiating a medical device correction involving the current labeling for this electronic product may be missing a required Certification Statement under 21 CFR Subchapter J (Radiological Health Regulations). Specifically, the following statement may have been omitted: Certification Statement: This electronic product complies with all applicable performance standards prescribed under 21 CFR Subchapter J, as enforced by the U.S. Food and Drug Administration at the date of manufacturing.

L
G
GRACE & FIRE PTY LTD — Ultra Violette Mini SPF Mates kit includes: Velvet Screen SPF 50 Blurring Mineral SkinScreen, Zinc Oxide 22.75%, 15 mL, 0.5 fl. oz. per tube, Supreme Sunscreen SPF 50, 15 mL 0.

Subpotent/Super-potent Product: Testing of the SPF 50 sunscreen revealed inconsistency in SPF results ranging from SPF 4, 10, 21, 26, 33, 60, 61, and 64.

G
GRACE & FIRE PTY LTD — Ultra Violette, Velvet Screen SPF 50, Blurring Mineral Skinscreen, Zinc Oxide 22.75%, a) 15 mL, 0.5 fl. oz. per tube, NDC 84803-106-01, UPC 9355909006068, b) 50 mL, 1.7 fl. oz.

Subpotent/Super-potent Product: Testing of the SPF 50 sunscreen revealed inconsistency in SPF results ranging from SPF 4, 10, 21, 26, 33, 60, 61, and 64.

G
GE Medical Systems, LLC — MAC VU360 Acquisition Trunk Cable and Module Holder, high performance, multichannel resting electrocardiograph, Model Number 2030360-001, Product Codes VU2 and VU4

If a user incorrectly places the Acquisition Module into the Acquisition Module Holder, it can result in excessive bending and wear of the Acquisition Trunk Cable. This is visible as a damaged sheath or exposed wires. If a damaged Acquisition Trunk Cable is not recognized and removed from service, the cable can overheat with continued use.

A
Abbott Point Of Care Inc. — i-STAT EG6+ cartridge. List Number: 03P77-25.

Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurement ranges for pH and PO2.

A
Abbott Point Of Care Inc. — i-STAT CG4+ cartridge (white). List Number: 03P85-25.

Lack of a 510(k) premarket clearance for the i-STAT CG4+ cartridge to reflect updated sample type for lactate and measurement ranges for pH and PO2.

M
Medline Industries, LP — Medline Kits: 1) VALVE PACK-LF, Model Number: DYNJ0415366P; 2) VALVE PACK-LF, Model Number: DYNJ0415366Q

The kits contain certain lots of cannula products where the catheter may not retain its shape.