Synthes (USA) Products LLC — LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibia

Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm.

I
Integra LifeSciences Corp. (NeuroSciences) — CODMAN Disposable Perforator 9mm. Cranial Perforator.

Expansion of recall of size 14 mm Codman Disposable Perforators for inadequate welds to size 11 mm and 9 mm Codman Disposable Perforators out of an abundance of caution.

D
DSAART LLC — Brand Name: AART Pectoral Implant Product Name: Pectoral Implant Model/Catalog Number: 500-106-5 Pectoral Implant style 1 size 6 left UDI-DI code: B4905001065 500-106-6 Pectoral Im

Due to manufactured products that deviated from established specifications and being distributed without obtaining the required FDA premarket approvals or clearances.

A
ASP Global, LLC. dba Anatomy Supply Partners, LLC. — SAFE-T-FILL Micro Collection Tube Extender, Model 09 1004

Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.

D
DSAART LLC — Brand Name: AART Gluteal Implant Product Name: Gluteal Implant Model/Catalog Numbers: 501-101 - Gluteal Implant style 1 size 1 UDI-DI code: B490501101 501-103 - Gluteal Implant sty

Due to manufactured products that deviated from established specifications and being distributed without obtaining the required FDA premarket approvals or clearances.

N
S
P
PHILIPS MEDICAL SYSTEMS — Model: Incisive CT for Brazil SKD; Product Code (REF): 728146; Software Versions: 5.1.0.X & 5.1.1.X;

Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation), the WED (Water Equivalent Diameter) value might set itself to zero, and this may lead to an insufficient dose setting after the surview. If operator misses the insufficient dose and the WED value in User Interface and continues with the subsequent helical/Axial scans, then the obtained images will be noisy due to low dose setting. Issue 3: Due to a software failure, the ECG wave file is not saved. As a result, the cardiac offline image reconstruction fails due to the missing ECG wave file, and the user might rescan the patient. Issue 4: This issue is software fault in which the patient orientation will be changed to NULL under specific random scenarios resulting in radiation being delivered from the wrong orientation (surview scan) or in the wrong position (clinical scan) and will generate incorrect and undiagnosable surview/clinical images. Patients may be rescanned for surview and/or clinical images. Issue 5: During scanning, the preview image display is not consistent, the images are not arriving at a fixed rate, skipping images from time to time during scanning. The obtained images might not support making the clinical diagnosis and the user might decide to re-scan the patient. Issue 6: If a user presses the left and middle/right mouse buttons together or afterwards in a short time, both commands are executed and an incorrect auto ROI is created. This ROI coordinates won t be assigned as the object is empty. As a result, the threshold of contrast level will not be reached in the UI and this will prevent the system from automatically triggering the subsequent clinical scan. There is potential safety risk identified for additionally tracker shots and/or rescan of a patient when the clinical scan is not triggered. This issue only affects the manual ROI process.

C
Covidien, LP — Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); BOX LOBECTOMIA VATS 3D; CFN: BOX06170V4;

A manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion.

P
PHILIPS MEDICAL SYSTEMS — Model: CT 5300; Product Code (REF): 728285; Software Versions: 5.1.0.X & 5.1.1.X;

Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation), the WED (Water Equivalent Diameter) value might set itself to zero, and this may lead to an insufficient dose setting after the surview. If operator misses the insufficient dose and the WED value in User Interface and continues with the subsequent helical/Axial scans, then the obtained images will be noisy due to low dose setting. Issue 3: Due to a software failure, the ECG wave file is not saved. As a result, the cardiac offline image reconstruction fails due to the missing ECG wave file, and the user might rescan the patient. Issue 4: This issue is software fault in which the patient orientation will be changed to NULL under specific random scenarios resulting in radiation being delivered from the wrong orientation (surview scan) or in the wrong position (clinical scan) and will generate incorrect and undiagnosable surview/clinical images. Patients may be rescanned for surview and/or clinical images. Issue 5: During scanning, the preview image display is not consistent, the images are not arriving at a fixed rate, skipping images from time to time during scanning. The obtained images might not support making the clinical diagnosis and the user might decide to re-scan the patient. Issue 6: If a user presses the left and middle/right mouse buttons together or afterwards in a short time, both commands are executed and an incorrect auto ROI is created. This ROI coordinates won t be assigned as the object is empty. As a result, the threshold of contrast level will not be reached in the UI and this will prevent the system from automatically triggering the subsequent clinical scan. There is potential safety risk identified for additionally tracker shots and/or rescan of a patient when the clinical scan is not triggered. This issue only affects the manual ROI process.

B
P
PHILIPS MEDICAL SYSTEMS — Model: Incisive CT; Product Code (REF): (1) 728143, (2) 728144; Software Versions: 5.1.0.X & 5.1.1.X;

Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation), the WED (Water Equivalent Diameter) value might set itself to zero, and this may lead to an insufficient dose setting after the surview. If operator misses the insufficient dose and the WED value in User Interface and continues with the subsequent helical/Axial scans, then the obtained images will be noisy due to low dose setting. Issue 3: Due to a software failure, the ECG wave file is not saved. As a result, the cardiac offline image reconstruction fails due to the missing ECG wave file, and the user might rescan the patient. Issue 4: This issue is software fault in which the patient orientation will be changed to NULL under specific random scenarios resulting in radiation being delivered from the wrong orientation (surview scan) or in the wrong position (clinical scan) and will generate incorrect and undiagnosable surview/clinical images. Patients may be rescanned for surview and/or clinical images. Issue 5: During scanning, the preview image display is not consistent, the images are not arriving at a fixed rate, skipping images from time to time during scanning. The obtained images might not support making the clinical diagnosis and the user might decide to re-scan the patient. Issue 6: If a user presses the left and middle/right mouse buttons together or afterwards in a short time, both commands are executed and an incorrect auto ROI is created. This ROI coordinates won t be assigned as the object is empty. As a result, the threshold of contrast level will not be reached in the UI and this will prevent the system from automatically triggering the subsequent clinical scan. There is potential safety risk identified for additionally tracker shots and/or rescan of a patient when the clinical scan is not triggered. This issue only affects the manual ROI process.

C
Covidien, LP — Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); CFN: SIGSDL45CTVT;

A manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion.

D
DJO, LLC — Cold Form Wrap: Back, REF: DP163CT03-BLK-L,XL, and Shoulder, REF: DP163CT02

The IFU (Information for Use) on the wrap instructs the user to heat the hot and cold compression wrap in the microwave for 3 minutes, which may cause the wrap to leak or break at the seams and/or reach temperatures capable of causing a second degree burn.

L
LeMaitre Vascular, Inc. — Artegraft¿ Collagen Vascular Graft; REF: AG1015;

The device was incorrectly packed in the wrong size labeled outer packaging.

S
Stryker Sustainability Solutions — HARMONIC FOCUS Shears + Adaptive Tissue Technology, 9 cm Length REF: HAR9F, Reprocessed Device for Single Use, STERILEEO, RX ONLY

Due to manufacturing issue, reprocessed shears, may become stuck in the "on position" resulting in continuous activation.

S
Southwind Foods LLC dba Great American Seafood Imports Co. — Tovala branded Frozen raw phosphate free shrimp; 5oz IQF bag UPC: 829944085788

Product manufactured under insanitary conditions whereby it may have been contaminated with Cesium-137 (Cs-137). This is an expansion to the original recall.

S
Southwind Foods LLC dba Great American Seafood Imports Co. — Arctic Shores; Best Yet; Great American; Master Catch branded Frozen Raw shrimp; 2lbs; 1lbs; UPC: 041130810923; 041512179471; 04113081097

Product manufactured under insanitary conditions whereby it may have been contaminated with Cesium-137 (Cs-137). This is an expansion to the original recall.

S
Southwind Foods LLC dba Great American Seafood Imports Co. — Arctic Shores branded Frozen cooked salad shrimp; 6 oz bag UPC: 041130811685

Product manufactured under insanitary conditions whereby it may have been contaminated with Cesium-137 (Cs-137). This is an expansion to the original recall.

S
Southwind Foods LLC dba Great American Seafood Imports Co. — Great American branded Frozen raw shrimp skewer; 1lb bag UPC: 829944002716; 829944002372

Product manufactured under insanitary conditions whereby it may have been contaminated with Cesium-137 (Cs-137). This is an expansion to the original recall.

S
Southwind Foods LLC dba Great American Seafood Imports Co. — Arctic Shores; Best Yet; Great American; Master Catch branded Frozen cooked shrimp; 2lbs; 1lbs; 12oz bags; UPC: 041130810497; 04113081041

Product manufactured under insanitary conditions whereby it may have been contaminated with Cesium-137 (Cs-137). This is an expansion to the original recall.

E
O
Optiwize Inc. — OptiWize Health OptiWize Collagen Plus 320 count bottle; White bottle with black lid.

The firm initiated a voluntary recall after NSF testing showed that labeled ingredient quantities did not align with analytical test results, manganese level was below label quantity.