Recall

All recalls7,869 total

F
Fresenius Medical Care Holdings, Inc. — Optiflux¿ High Flux E-beam Dialyzer Model Numbers: (1) 0500316E OPTIFLUX F160NRe 12/CS 1.5SA EBEAM, (2) 0500318E OPTIFLUX F180NRe 12/CS 1.7SA EBEAM, (3) 050032

To provide further information to the user to ensure proper use of the medical device. the caps had been changed from a threaded version to a press-on version whereby the user must press the caps firmly onto the dialyzer to securely affix the cap.

S
Siemens Medical Solutions USA, Inc — MAGNETOM Skyra (DE). Model Number: 10432915. 510(k) Numbers: K101347, K111242, K123510, K133435, K140253, K141977, K142515, K153343, K173592, K202014, K231560.

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

S
Siemens Medical Solutions USA, Inc — Biograph mMR. Model Number: 10433372.

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

S
Siemens Medical Solutions USA, Inc — MAGNETOM Verio. Model Number: 10276755.

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

S
Siemens Medical Solutions USA, Inc — MAGNETOM Lumina (DE). Model Number: 11344916.

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

H
Howmedica Osteonics Corp. — 1. EXETER V40 STEM 44MM NO 2, Model/Catalog Number: 0580-1-442; 2. EXETER V40 STEM 37.5MM NO 0, Model/Catalog Number: 0580-1-352;

a potential product mix where the package labeled as Catalog Number 0580-1-442, lot A00976, may contain Catalog Number 0580-1-352, lot G8754849, and vice versa. Patient labels may also be impacted.

S
Siemens Medical Solutions USA, Inc — MAGNETOM Spectra as the following: 1. MAGNETOM Spectra. Model Number: 10837643. 2. MAGNETOM Spectra (CN). Model Number: 10655588.

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

S
Siemens Medical Solutions USA, Inc — MAGNETOM Skyra Fit BioMatrix. Model Number: 11516217.

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

S
Siemens Medical Solutions USA, Inc — MAGNETOM Vida (DE). Model Number: 11060815. 510(k): K170396, K173617, K181433, K183254, K192924, K203443, K213693, K213805, K220425, K231560.

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

S
Siemens Medical Solutions USA, Inc — MAGNETOM Vida Fit. Model Number: 11410481.

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

S
Siemens Medical Solutions USA, Inc — MAGNETOM Skyra fit. Model Number: 10849580.

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

S
Siemens Medical Solutions USA, Inc — MAGNETOM Cima.X (DE). Model Number: 11647158.

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

S
Siemens Medical Solutions USA, Inc — MAGNETOM Verio Dot Upgrade. Model Number: 10684334.

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

B
Boston Scientific Corporation — Extractor Pro RX Retrieval Balloon Catheter; UPN (REF): (1) M00547030 (9-12mm Below), (2) M00547000 (9-12mm Above);

The product in incorrectly labeled. The label indicates that the skive hole should be above the balloon, while it is actually positioned below, and vice versa.

S
Siemens Medical Solutions USA, Inc — MAGNETOM Connectom.X. Model Number: 11371480.

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

S
Siemens Medical Solutions USA, Inc — MAGNETOM Prisma. Model Number: 10849582.

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

S
Siemens Medical Solutions USA, Inc — BIOGRAPH One (DE). Model Number: 11689172.

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

S
Siemens Medical Solutions USA, Inc — MAGNETOM Verio Dot. Model Number: 10684333.

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

D
Dexcom, Inc. — Brand Name: Dexcom G6 Continuous Glucose Monitoring System Product Name: Dexcom G6 and G6 Pro Android US CGM App Model/Catalog Number: SW11678 Software Version: version 1.15.0 Produ

A software defect in version v1.15.0 of the G6 Android app can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glucose values, alarms, alerts or notifications. This could result in the missed detection of a hyperglycemic or hypoglycemic event, protentional resulting in severe hyperglycemia, diabetic ketoacidosis (DKA), or hyperosmolar hyperglycemic state (HHS).

V
Vantive US Healthcare LLC — Sharesource Connectivity Platform for Use with Homechoice Claria Product Code 5CGM01 software versions 7.13.3 (China), 8.13.2 (APAC and Korea), and 8.13.4 (US, EMEA, an

Vantive has identified a software defect within the Sharesource Claria software, which may cause a patient s updated prescription program settings to not be properly saved on the actual cycler that the patient is using to perform their therapy. As a result, the clinician may think that the patient is receiving the correct prescribed therapy because of what is erroneously displayed on the clinician's remote Sharesource application, when in fact the cycler that is with the patient is actually delivering an outdated and incorrect therapy.

A
Amerisource Health Services LLC — chlorproMAZINE Hydrochloride Tablets, USP, 200 mg, 100 (10x10) blisterpack per carton, Rx Only, Distributed by: American Health Packaging, Columbus, Ohio 43217. NDC#:

Presence of a foreign substance.A specific lot of auxiliary polyester coil, used in product packaging at the manufacturing site, was detected with presence of a micro-organism. No micro-organism was detected on any tablets.

A
Amerisource Health Services LLC — chlorproMAZINE Hydrochloride Tablets, USP, 50 mg, 100 (10x10) blisterpacks, Rx Only, Distributed by: American Health Packaging, Columbus, Ohio 43217. 50 Tablet Carton

Presence of a foreign substance.A specific lot of auxiliary polyester coil, used in product packaging at the manufacturing site, was detected with presence of a micro-organism. No micro-organism was detected on any tablets.

A
Amerisource Health Services LLC — chlorproMAZINE Hydrochloride Tablets, USP, 100 mg, 100 (10x10) blisterpack per carton, Rx Only, Distributed by: American Health Packaging, Columbus, Ohio 43217. Cart

Presence of a foreign substance.A specific lot of auxiliary polyester coil, used in product packaging at the manufacturing site, was detected with presence of a micro-organism. No micro-organism was detected on any tablets.

Z
Zydus Pharmaceuticals (USA) Inc — Succinylcholine Chloride Injection, USP 200 mg/10 mL (20 mg/mL), Rx only, 10 mL Multiple-dose Vial, Sterile, Manufactured by Zydus Lifesciences Ltd, Vadodara, India;

Failed Impurities/Degradation Specifications: Out-of-Specification test results obtained for any individual unknown degradation impurity.

A
Amerisource Health Services LLC — chlorproMAZINE Hydrochloride Tablets, USP, 25 mg, packaged as a) 50 (5x10) blisterpacks, NDC: 60687-430-65; (Individual Dose NDC: 60687-430-11); b) 100 (10x10) bliste

Presence of a foreign substance.A specific lot of auxiliary polyester coil, used in product packaging at the manufacturing site, was detected with presence of a micro-organism. No micro-organism was detected on any tablets.

U
Unichem Pharmaceuticals USA Inc. — Cyclobenzaprine Hydrochloride Tablets, USP, 10 mg, 90-count bottle, RX only, Manufactured by: Unichem Laboratories, Ltd, Pilerne Ind. Estate, Pilerne, Bardez, Goa

Labeling: Label Mix Up; Bottles of Meloxicam USP, 7.5mg, 90-count tablets (yellow in color), were labeled as Cyclobenzaprine Hydrochloride Tablets USP, 10 mg 90-count tablets (blue in color).

← Previous

Page 87 of 158

Next →