Medline Industries, LP — Medline medical convenience kits packaged as: STERILE RESUTURE PACK, Kit SKU DYNDL1980A

Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.

M
Medline Industries, LP — Medline medical convenience kits packaged as: 1) CANCER SERVICE PORT KIT, Kit SKU DT22710; 2) LINE KIT, Kit SKU DYNDA1857A; 3) PICC REMOVAL KIT, Kit SKU DYNDC1978A; 4

Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.

M
Medline Industries, LP — 1) UNIVERSAL KIT, Kit SKU DT22010B; 2) VAD ACCESS KIT, Kit SKU DT22790C; 3) VAD ACCESS KIT, Kit SKU DT22790C; 4) PROCEDURE TRAY, Kit SKU DYNDA2966A; 5) PROCEDURE TRAY,

Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.

M
Medline Industries, LP — Medline medical convenience kits packaged as: BLOOD CULTURE KIT NS, Kit SKU DYNDH2088

Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.

M
Medline Industries, LP — Medline medical convenience kits packaged as: ADULT CATH SECUREMENT DRSG, Kit SKU DYNDC3109

Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.

M
Medline Industries, LP — Medline medical convenience kits packaged as: MERCY CIRCUMCISION TRAY, Kit SKU CIT6135B

Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.

M
Medline Industries, LP — Medline medical convenience kits packaged as: 1) POWERGLIDE DRESSING CHANGE KIT, Kit SKU DYNDH1576; 2) CENTRAL LINE TRAY W/ CHLRP SWB, Kit SKU RWM-DC57KC.

Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.

M
Medline Industries, LP — Medline 14 Fr Suction Catheter Kits SKU DYND40972 - respiratory convenience kit

A smaller open suction catheter, the 8 Fr Suction Catheter, may have been packed within the kit instead of the required 14 Fr Suction Catheter. Use of the smaller size catheter may result in hypoxia and/or respiratory failure related to partial or full airway obstruction. Airway obstruction may require immediate emergent medical intervention to prevent respiratory failure or death. Other potential outcomes include prolonged procedure, airway irritation, or prolonged dyspnea.

Z
Zimmer Surgical Inc — Brand Name: Zimmer Tourniquet Systems Product Name: A.T.S.¿ 3200TS Tourniquet Systems Model/Catalog Number: 60320010100 Software Version: Software version v2.06 or prior are w

Device user interface (screen) may freeze or become unresponsive to touch when operating in certain non-English language settings. Issue may cause delay or extension while alternate device is located.

M
Medline Industries, LP — Medline medical convenience kits packaged as: PEDIATRIC CATH SECUREMENT DRSG, Kit SKU DYNDC3108

Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.

G
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING — GE Healthcare Omni Legend, sold under the following names: OMNI Legend 16, Omni Legend 21, OMNI Legend 32, Omni Legend PET Gantry 16cm - MEA, Omni Le

There is a potential intermittent issue on certain Omni Legend systems that can result in a streaking artifact in the PET clinical scan images. This streaking artifact is most easily identified in transaxial slices on the acquisition console (both corrected and non-corrected for attenuation).

Z
Zimmer Surgical Inc — Brand Name: Zimmer Tourniquet Systems Product Name: A.T.S.¿ 5000TS Tourniquet Systems Model/Catalog Number: 60500010100 Software Version: Software version v2.04 or prior are w

Device user interface (screen) may freeze or become unresponsive to touch when operating in certain non-English language settings. Issue may cause delay or extension while alternate device is located.

B
Beckman Coulter, Inc. — CellMek SPS Sample Preparation System, REF: C44603, with software

Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.

F
Focalyx Technologies, LLC. — Focalyx Fusion

Software device that is an accessory for image-guided interventional and diagnostic procedures involving prostate has accessories that may not function properly with Windows 10, which may cause patient harm during biopsy or cause treatment/therapy delays, so firm is recommending stop use until device is verified/validated with Windows 11.

B
M
Medline Industries, LP — Reprocessed St. Jude Supreme and Response Diagnostic Electrophysiology Catheters: St. Jude Medical Supreme Diagnostic Electrophysiology Catheter 4 Elec. 2-5-2mm 401211RH

These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025

P
Philips North America — Patient Information Center iX; Software Version Number: 4.5.0

A recent software (SW) patch modifies Mobile Event Notification filter settings without providing any indication to the user when upgrading the system.

M
Medline Industries, LP — Reprocessed Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheters: Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheter Use on GE Systems 10439072RH;

These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025

M
Medline Industries, LP — Reprocessed Livewire Steerable Diagnostic Electrophysiology Catheters: Livewire Steerable Electrophysiology Catheter,10 Elec. 2-5-2mm 401582RH 401940RH; Livewire Steer

These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025

M
Medline Industries, LP — Medline ReNewal Reprocessed ViewFlex Xtra ICE Catheters: ViewFlex Xtra ICE Catheter 9F x 90cm D087031RH 20197344019554¿

These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025

L
LSL Healthcare Inc. — LSL Healthcare, Central Line Dressing Kit, Model/Catalog Number: 2519CP, medical procedure convenience kit

BD ChloraPrep Triple Swabsticks, a component within the kits, exhibit an open seal on the packaging of the applicators.

M
Medline Industries, LP — Reprocessed Abbott Inquiry Steerable Diagnostic Catheters: Abbott Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81105RH 81172RH; Abbott Inquiry Steerable Di

These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025

M
Medline Industries, LP — Medline ReNewal Reprocessed Webster CS Catheters: Webster CS Catheter 10 Electrodes 2-8-2mm Spacing w/AutoID w/ EZ Steer BD710DF282CRH; Webster CS Catheter 10 Electrodes,

These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025

M
Medline Industries, LP — Reprocessed Siemens ACUSON AcuNav Diagnostic Ultrasound Catheters: ACUSON AcuNav Ultrasound Catheter, Use on GE Systems 10135910RH; 10043342RH ACUSON AcuNav Ultrasound

These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025

O
Z
T
Tornier, Inc — TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM NON STERILE, catalog number DWJ318

A specific lot of screws is incorrectly labeled. Screws are labeled as 5.0mmx18mm but actual screws are 5.00mmx14mm. If mismatch is detected during surgery, the issue may cause a minor delay in surgery. If not detected and incorrect screw is used, issue may increase risk of earlier mechanical failure, loosening, and need for revision surgery.

I
ICU Medical Inc. — Portex Spinal Tray, Item No. A3729-24 A3595

Affected lots of syringes may have a crack on the syringe collar, which may lead to delayed or repeated blood draw/fluid delivery. This issue may also lead to blood leakage during sampling of arterial blood or leakage during fluid delivery.

S
Siemens Medical Solutions USA, Inc — NAEOTOM Alpha Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPECTS

To remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance.

C
Canon Medical System, USA, INC. — Alphenix INFX-8000H, interventional fluoroscopic x-ray system

It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed

B
Boston Scientific Corporation — *HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 8mm x 8mm; Material Number (UPN): M00553530;

Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.

C
Canon Medical System, USA, INC. — Alphenix INFX-8000V, interventional fluoroscopic x-ray system

It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed

B
Boston Scientific Corporation — AXIOS Stent and Electrocautery-Enhanced Delivery System - 8mm x 8mm; Material Number (UPN): M00553690;

Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.

C
Canon Medical System, USA, INC. — Alphenix INFX-8000C, interventional fluoroscopic x-ray system

It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed