Recall

All recalls7,815 total

M
Medline Industries, LP — Medline Convenience Kits: 1) KIT TRIPLE LUMEN INSERTION, Model Number: DYNDM1031A

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

M
Mint Medical GmbH — Mint Lesion; Software Versions: 3.4.0 up to 3.9.5.;

If the connection between a mint Lesion workstation and the mint Lesion server is interrupted while performing a read, in rare cases some information may be lost or incorrectly linked to other patients.

M
Medline Industries, LP — Medline Convenience Kits: 1) CHEST TUBE INSERTION TRAY, Model Number: DYNDA2941A; 2) CHEST/BREAST-LF, Model Number: DYNJ83950; 3) HERNIA REPAIR TOTE, Model Number: DYNJ90

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

M
Medline Industries, LP — Medline Convenience Kits: 1) KIT SURG ONC EXCISION BREAST, Model Number: DYKMBNDL84A; 2) KIT SURG ONC EXPLORATORY LAP, Model Number: DYKMBNDL87A; 3) CT BIOPSY TRAY, Model

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

M
Medline Industries, LP — Medline Convenience Kits: 1) KIT TRIPLE LUMEN INSERTION, Model Number: DYNDM1031A

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

P
Philips North America Llc — Brilliance iCT; Product Code (REF): 728306;

As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.

M
Medline Industries, LP — Medline Convenience Kits: 1) KIT ACES INSERT PEG, Model Number: DYKMBNDL130; 2) KIT ACES INSERT PEG, Model Number: DYKMBNDL130A; 3) KIT ACES INSERT PERCUTANEOUS T, Model

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

M
Medline Industries, LP — Medline Convenience Kits: 1) DRAPE PACK-CHOICE, Model Number: DYNJ63118A

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

P
Philips North America Llc — Spectral CT; Product Code (REF): 728333;

As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.

M
Medline Industries, LP — Medline Convenience Kits: 1) LVAD DRIVELINE TRAY, Model Number: DM1035; 2) KIT SURG ONC INSERT CENT.VENOU, Model Number: DYKMBNDL43

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

M
Medline Industries, LP — Medline Convenience Kits: 1) ADMIT KIT, Model Number: DYKA1343A

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

M
Medline Industries, LP — Medline Convenience Kits: 1) BAP NORTH ROBOTIC SURGERY CDS, Model Number: CDS982407N; 2) TOT/ GYN LITHOTOMY CDS, Model Number: CDS985431K; 3) TOT/ GYN LITHOTOMY CDS, Mode

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

M
Medline Industries, LP — Medline Convenience Kits: 1) MAJOR LAPAROTOMY CDS, Model Number: CDS860015U; 2) MAJOR LAPAROTOMY CDS, Model Number: CDS860015V; 3) MAJOR LAPAROTOMY CDS, Model Number: CDS

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

M
Medline Industries, LP — Medline Convenience Kits: 1) KIT GENERAL CLOSURE, Model Number: DYKMBNDL200A

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

M
Medline Industries, LP — Medline Convenience Kits: 1) KIT STC TRACHEOSTOMY, Model Number: DYKMBNDL55; 2) KIT STC TRACHEOSTOMY, Model Number: DYKMBNDL55A; 3) TRACH TOTE, Model Number: DYNJ85691

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

M
Medline Industries, LP — Medline Convenience Kits: 1) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069G; 2) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069I; 3) HEAD & NECK CDS-LF, Model

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

M
Medline Industries, LP — Medline Convenience Kits: 1) LITHOTOMY PACK, Model Number: DYNJ83185; 2) LITHOTOMY PACK, Model Number: DYNJ83185A;

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

M
Medline Industries, LP — Medline Convenience Kits: 1) PPE KIT, Model Number: DYK1011945P

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

M
Medline Industries, LP — Medline Convenience Kits: 1) LVAD DRIVELINE TRAY, Model Number: DM1035A; 2) LVAD DRIVELINE TRAY, Model Number: DM1035; 3) CENTRAL LINE DRESSING CHANGE TRAY, Model Number:

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

M
Medline Industries, LP — Medline Convenience Kits: 1) ANCILLARY PACK-LF, Model Number: DYNJ0185290G; 2) OPEN HEART ANESTHESIA ICC-A, Model Number: DYNJ905567A

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

M
Medline Industries, LP — Medline Convenience Kits: 1) WMC I D PACK-LF, Model Number: DYNJ50806L

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

M
Medline Industries, LP — Medline Convenience Kits: 1) ROBOTIC GYN CDS-1, Model Number: CDS982893I; 2) GYN LAPAROSCOPY CDS, Model Number: CDS983300J; 3) NHP ROBOTIC GYNE CDS, Model Number: CDS9842

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

P
Philips North America Llc — IQon Spectral CT; Product Code (REF): 728332;

As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.

M
Medline Industries, LP — Medline Convenience Kits: 1) BURN PACK, Model Number: DYNJ15668R; 2) BURN PACK-LF, Model Number: DYNJ42829B; 3) BURN CARE PACK-LF, Model Number: DYNJT3437

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

M
Medline Industries, LP — Medline Convenience Kits: 1) BAPTIST FLOYD ENDO GI KIT, Model Number: DYKE1455B; 2) GI LAB OTHER ENDO KIT, Model Number: DYKE1721

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

M
Medline Industries, LP — Medline Convenience Kits: 1) PLASTIC BIN W/MISC KITS, Model Number: ACC010499; 2) KIT UROLOGY URETHROPLASTY, Model Number: DYKMBNDL117C; 3) KIT PLASTIC INSERT TISSUE EXPA

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

M
Medline Industries, LP — Medline Convenience Kits: 1) NEURO SHUNT- CODMAN, Model Number: CDS860018G; 2) NEURO SHUNT- CODMAN, Model Number: CDS860018I; 3) KIT NEUR FUS LUM POST CRANDALL, Model Num

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

M
Medline Industries, LP — Medline Convenience Kits: 1) ENDO KIT W/SYRINGE, Model Number: DYKE1743; 2) BGMC ANGIOGRAPHY PACK-LF, Model Number: DYNJ0800934D; 3) GENERAL ENDO PACK-LF, Model Number: D

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

M
Medline Industries, LP — Medline Surgical Gowns: 1) GOWN,NONRNF,L,30/CS, Model Number: DYNJP2001; 2) GOWN,SIRUS,NONRNF,SETINSLV,L,20/CS, Model Number: DYNJP2001S; 3) GOWN,ECLIPSE,NONRNF,XL,ST,30/

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

M
Medline Industries, LP — Medline Convenience Kits: 1) OB KIT, Model Number: DYKM2326; 2) KIT LABOR DELIVERY TUBAL LIGAT, Model Number: DYKMBNDL165; 3) KIT LABOR & DELIVERY D&C/D&E, Model Number:

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

← Previous

Page 46 of 157

Next →