L
Lupin Pharmaceuticals Inc. — clomiPRAMINE Hydrochloride Capsules USP 25 mg, 100 count bottles, 5801 Pelican Bay Boulevard, Suite 500, Naples, Florida 34108, NDC# 68180-492-01

Failed Impurities/Degradation Specifications: an out of specification result observed in degradation product test (any unspecified degradation product) during retention sample testing at expiry.

M
Microbiologics Inc — LYFO-DISK, M. Canis Derived from ATCC 36299, packaged as: 1) REF 0894K (6-pack), 2) REF 0894P (2-pack)

KWIK-STIK REF numbers: 0894P, 0894K and LYFO DISK" 0894L may result in failure to recover the target microorganism. All lots of REF 894-76 are impacted.

S
Spiggle & Theis Mt Gmbh — Injection needle, bayonet, Luer-Lock, 1. Model no. 50-353-23, working length 25 cm, total length 31cm, nonsterile; 2. Model no. 50-345-23, working length 24 cm, total l

Affected product show elevated cytotoxicity values. Use of affected injection needles may present a risk of intolerance reaction or allergic reaction.

M
Maquet Cardiopulmonary Gmbh — Sprinter Cart/Sprinter Cart XL Infusion Pole. Component used to hold 2 x 2 kg infusion bags.

Device does not comply with the labeling requirements of international standard IEC 60601-1.

M
Microbiologics Inc — LYFO-DISK, M. Canis Derived from ATCC 36299, packaged as: 1) REF 0894L

KWIK-STIK REF numbers: 0894P, 0894K and LYFO DISK" 0894L may result in failure to recover the target microorganism. All lots of REF 894-76 are impacted.

B
Beckman Coulter Inc. — UIBC (Unsaturated Iron Binding Capacity), REF: OSR61205,

Beckman Coulter identified that the Unsaturated Iron Binding Capacity (UIBC) assay is not meeting their labeled hemoglobin interference claims of no significant interference (i.e., less than 10%) up to 200 mg/dL hemoglobin as indicated in their instructions for use. Initial internal testing confirmed that samples with low UIBC levels failed the claimed hemolysis interference specifications with reported biases up to -43.6 % UIBC when hemolyzed serum samples contained 200 mg/dL hemoglobin.

S
Siemens Healthcare Diagnostics Inc — RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.

Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and their derivatives. Four complaints were filed related to this event that led to Siemens Healthcare Diagnostics identifying an error in the manufacturing process and initiated corrective actions. Use of the affected device may result in errors during QC which could lead to a delay in patient treatment. If bias generated from this issue is not caught during QC erroneous results may cause temporary or medically reversible adverse health consequences to patients.

W
H
Hamilton Medical AG — Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.

Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete rebreathing of exhaled gases, which could lead to acute hypercapnia, respiratory acidosis, and if unrecognized organ dysfunction.

B
bioMerieux, Inc. — VITEK¿ 2 Gram-negative AST Test Kits (Cards) containing Colistin formulation cs02n. VITEK¿ 2 is an automated system consisting of instruments, software and reagent cards designed

Potential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex used in specific cards

B
Beckman Coulter, Inc. — DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457

Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.

B
Beckman Coulter, Inc. — DxI 600 Access Immunoassay Analyzer W/Spot B, Part Number A71460

Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.

B
Beckman Coulter, Inc. — DxI 800 Access Immunoassay Analyzer W/Spot B, Part Number A71456

Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.

C
C SWEET PASTRY LLC — C Sweet Ghouribeh with Pistachios 7 oz., UPC 852664042755

Undeclared soy, contains statement is present but does not list pistachios

C
C SWEET PASTRY LLC — C Sweet Maamoul with Pistachios 8 oz., UPC 721782348967

Undeclared soy, contains statement is present but does not list pistachios

B
Boston Scientific Corporation — FARAWAVE 1.0 Pulsed Field Ablation Catheter, Product ID M004PF41M401

The potentially impacted units were manufactured using specific equipment that may have caused cracks in the electrode bands on the catheter's distal end.

B
Beckman Coulter, Inc. — MicroScan Neg MIC 3J REF C54814

Due to a manufacturing issue - missing a drug in well 4/4 but contain double the amount of antibiotics

B
Beckman Coulter, Inc. — DxI 600 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71461

Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.

N
P
Preat Corp — NobelBiocare Tri-Lobe-compatible WP X 6.0mm Engaging Titanium Base, REF: 9000662

Engaging titanium base was manufactured 180 degrees around the post from where it should be located in relation to the connection geometry; if the flat is on the wrong side of the post, the crown will not fit onto the abutment in the correct orientation, resulting in restoration delay.

B
Beckman Coulter, Inc. — UniCel DxI 600 Access Immunoassay Analyzer, Part Number A30260

Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.

B
Beckman Coulter, Inc. — UniCel DxI 800 Access Immunoassay Analyzer, Part Number 973100

Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.

N
M
Moon N Sea GA LLC — Blue Fin Tuna, Net WT (variable weight) 50 lb brown corrugated carboard box. Boxes contain ice for temperature control. Keep Refrigerated.

RTE fish (sushi-salmon, tuna) processed under insanitary conditions. The firm was notified by the Georgia Department of Agriculture that the products were manufactured under unsanitary conditions.

E
Enjoy Nutrition, LLC; dba Weyland Brain Nutrition — Lithium Orotate, 5MG, Distributed by Enjoy Nutrition, LLC. UPC#: 702730893968, 702730894064

Under potency / misbranding: product fails to meet labeled lithium orotate potency. (Downstream recall) from supplier of ingredient.

M
Moon N Sea GA LLC — Yellow Fin Tuna, Net WT (variable weight) 50 lb brown corrugated carboard box. Boxes contain ice for temperature control. Keep Refrigerated.

RTE fish (sushi-salmon, tuna) processed under insanitary conditions. The firm was notified by the Georgia Department of Agriculture that the products were manufactured under unsanitary conditions.

E
Enjoy Nutrition, LLC; dba Weyland Brain Nutrition — Lithium Orotate, 10MG, Distributed by Enjoy Nutrition, LLC. UPC#: 702730894033, 702730894057

Under potency / misbranding: product fails to meet labeled lithium orotate potency. (Downstream recall) from supplier of ingredient.

M
Moon N Sea GA LLC — Fresh Scottish Salmon Fillet, Net WT (variable weight) 50 lb brown corrugated carboard box (up to 6 fillets per box). Boxes contain ice for temperature control. Keep Refrigerated.

RTE fish (sushi-salmon, tuna) processed under insanitary conditions. The firm was notified by the Georgia Department of Agriculture that the products were manufactured under unsanitary conditions.

M
Moon N Sea GA LLC — Fresh Chilean Salmon Fillet, Net WT (variable weight) 50 lb brown corrugated carboard box (up to 6 fillets per box). Boxes contain ice for temperature control. Keep Refrigerated.

RTE fish (sushi-salmon, tuna) processed under insanitary conditions. The firm was notified by the Georgia Department of Agriculture that the products were manufactured under unsanitary conditions.

E
Enjoy Nutrition, LLC; dba Weyland Brain Nutrition — Lithium Orotate, 20MG, Distributed by Enjoy Nutrition, LLC. UPC#: none

Under potency / misbranding: product fails to meet labeled lithium orotate potency. (Downstream recall) from supplier of ingredient.

E
Enjoy Nutrition, LLC; dba Weyland Brain Nutrition — Lithium Orotate, 2.5MG, Distributed by Enjoy Nutrition, LLC. UPC#: 719318414002; 709750005289

Under potency / misbranding: product fails to meet labeled lithium orotate potency. (Downstream recall) from supplier of ingredient.

E
Enjoy Nutrition, LLC; dba Weyland Brain Nutrition — Lithium Orotate, 1MG, Distributed by Enjoy Nutrition, LLC. UPC#: 717340952714

Under potency / misbranding: product fails to meet labeled lithium orotate potency. (Downstream recall) from supplier of ingredient.

G
Granules Pharmaceuticals Inc. — Metoprolol Succinate Extended-Release Tablets, USP, 25 mg, Packaged in a) 100-count bottle, NDC 70010-780-01; b) 500-count bottle, NDC 70010-780-05; Rx only, Manufactur

Failed Dissolution Specifications: Product failed to meet dissolution acceptance criteria in the stability studies at the 6th month (25¿C/60% RH) long-term.

O
Orthofix U.S. LLC — PILLAR SA Ti, STERILE, 37mm (W) x 28mm (D) x 18mm (H), 20-degrees, REF: 82-3713SP

A 37 mm implant width intervertebral body fusion device has a labeling discrepancy with incorrect width of 33 mm stated on the outer label side flap and inner label side flap information box. The 37 mm width on the front labels is correct.

P
Preat Corp — Preat, Astra EV-compatible 4.2mm Digital Analog REF: 9006603 (single) and 9006603-10(10 pack), Rx Only

Due to manufacturing error, digital analog rotational feature is 180 degrees off.