Recall
FDAHealthJune 26, 2025

Siemens Healthcare Diagnostics Inc — RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.

⚠️ Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and their derivatives. Four complaints were filed related to this event that led to Siemens Healthcare Diagnostics identifying an error in the manufacturing process and initiated corrective actions. Use of the affected device may result in errors during QC which could lead to a delay in patient treatment. If bias generated from this issue is not caught during QC erroneous results may cause temporary or medically reversible adverse health consequences to patients.

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
June 26, 2025

worldwide distribution - US Nationwide and the countries of Canada, Chile, Japan, Paraguay.

What's affected?

Product
RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.
Lot code
WW/08925

Additional details

Material Number: 10329097. GTIN Number: 00630414482163. Lot Number: WW/08925. Expiration Date: 12/03/2025.

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 1,477 units.

View original FDA notice →

Get Health recall alerts

Delivered to your inbox. Free. No account required.