Siemens Healthcare Diagnostics Inc — RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.
⚠️ Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and their derivatives. Four complaints were filed related to this event that led to Siemens Healthcare Diagnostics identifying an error in the manufacturing process and initiated corrective actions. Use of the affected device may result in errors during QC which could lead to a delay in patient treatment. If bias generated from this issue is not caught during QC erroneous results may cause temporary or medically reversible adverse health consequences to patients.
Is this relevant to me?
- Geography
- Nationwide
- Source agency
- FDA
- Category
- Health
- Published
- June 26, 2025
worldwide distribution - US Nationwide and the countries of Canada, Chile, Japan, Paraguay.
What's affected?
- Product
- RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.
- Lot code
- WW/08925
Additional details
Material Number: 10329097. GTIN Number: 00630414482163. Lot Number: WW/08925. Expiration Date: 12/03/2025.
What should I do?
Recommended action
Class II recall. Voluntary: Firm initiated Quantity affected: 1,477 units.