Recall
FDAHealthJune 26, 2026

Erbe USA Inc — ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000

⚠️ A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
June 26, 2026

US Nationwide distribution including in the states and territories of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming.

What's affected?

Product
ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000
Lot code
UDI-DI 04050147013797/ Serial Numbers: All serial numbers up to 11608268 Software Version: 1.0.3

Additional details

UDI-DI 04050147013797/ Serial Numbers: All serial numbers up to 11608268 Software Version: 1.0.3

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 11226 units.

View original FDA notice →