Recall
FDAHealthJune 15, 2026

TISSIUM SA — Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO4010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair dev

⚠️ Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
June 15, 2026

US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA.

What's affected?

Product
Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO4010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No
Lot code
Model

Additional details

Lot Code: Model No NCOCO4010; UDI-DI 03760279380148; Lot Numbers 251000041, 250700048, 260300046, 260400050, and 260300047

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated

View original FDA notice →