Boston Scientific Neuromodulation Corporation — INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part

⚠️ Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, including lead explant and replacement.

Is this relevant to me?

Geography
US, NY
Source agency
FDA
Category
Health
Published
June 17, 2026

US: NY. OUS: Canada, Portugal, Spain, France, Netherlands, Great Britain, Italy, Israel, Serbia, Greece, Germany, United Arab Emirates, Ireland, Hungary, Czech Republic, Brazil, Colombia, Argentina

What's affected?

Product
INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part of the Boston Scientific Spinal Cord Stimulator (SCS) Systems.
Lot code
Model/UDI-DI: SC-2317-50/08714729861614, SC-2317-70/08714729861638. All unused products.

Additional details

Model/UDI-DI: SC-2317-50/08714729861614, SC-2317-70/08714729861638. All unused products.

What should I do?

Recommended action

Class I recall. Voluntary: Firm initiated Quantity affected: 1079.