Recall
FDAHealthJuly 16, 2026

Biofire Defense — FilmArray NGDS Warrior Panel REF NGDS-ASY-0007 UDI-DI code: (01)00851458005136 The FilmArray NGDS Warrior Panel is a qualitative, multiplexed, nucleic acid-based in vitro diagnos

⚠️ Firm has identified mismatches between the outbreak strain and the reference strain used in the development of its qualitative, multiplexed, nucleic acid-based in vitro diagnostic panel test.

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
July 16, 2026

U.S. Nationwide distribution in the states of AL, CA, DC, FL, GA, HI, KY, LA, MA, MD, MN, NC, NH, NV, NY, OH, RI, TN, TX, UT, VA and WA.

What's affected?

Product
FilmArray NGDS Warrior Panel REF NGDS-ASY-0007 UDI-DI code: (01)00851458005136 The FilmArray NGDS Warrior Panel is a qualitative, multiplexed, nucleic acid-based in vitro diagnostic test intended for use with the FilmArray 2.0 system. The FilmArray NGDS Warrior Panel detects and identifies Bacillus anthracis, Yersinia pestis, Francisella tularensis, Coxiella burnetii, Ebola virus, and Marburg virus nucleic acids directly from human whole blood (EDTA).
Lot code
311825D

Additional details

Part Number: NGDS-ASY-0007 UDI-DI code: (01)00851458005136 Lot Numbers: 311825D 312025D 313526D 322925D

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 122 kits (3,660 pouches).