Innovasis, Inc — Anti-Torque, Vector, Part: LS-749 part of Vector Pedicle Screw System

⚠️ Pedicle screw system intended to provide immobilization/stabilization of spinal segments as adjunct to thoracic/lumbar/sacral spine fusion may have anti-torque instrument with undersized condition which could prevent mating tip from engaging with the pedicle screw tulip and rod during tightening, which may lead to surgical delay, intervention/alternative methods to perform final tightening step.

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
May 22, 2026

US Nationwide distribution in the states of TX, CA.

What's affected?

Product
Anti-Torque, Vector, Part: LS-749 part of Vector Pedicle Screw System
Lot code
LS-749: UDI-DI: M711BTA2020, M711LS7490, Serial Numbers: 002, 003, 007, 012, 014

Additional details

LS-749: UDI-DI: M711BTA2020, M711LS7490, Serial Numbers: 002, 003, 007, 012, 014

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 5.