Recall
FDAHealthMay 8, 2026

Instrumentation Laboratory — Brand Name: GEM Premier 7000 with iQM3 Product Name: GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Intelligent Quality Management 3 (iQM3) is used as the q

⚠️ Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
May 8, 2026

US Nationwide distribution.

What's affected?

Product
Brand Name: GEM Premier 7000 with iQM3 Product Name: GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Intelligent Quality Management 3 (iQM3) is used as the quality control (QC) and assessment system for the GEM Premier 7000. portable blood analyzer
Lot code
GEM Premier 7000: Part number: 00000015279 UDI: 0195226000924 Units distributed in US: GEM Premier 7000 Serial Nos. 26012062, 26012065, 26012069 and 26012070. Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634

Additional details

GEM Premier 7000: Part number: 00000015279 UDI: 0195226000924 Units distributed in US: GEM Premier 7000 Serial Nos. 26012062, 26012065, 26012069 and 26012070. Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 4 US.