Recall
FDAHealthJune 11, 2026

Boston Scientific Corporation — FARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PF21M402, IFU Reference Nu

⚠️ Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
June 11, 2026

Global Distribution. US Nationwide.

What's affected?

Product
FARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PF21M402, IFU Reference Number 51654495-01
Lot code
CL12695

Additional details

UDI-DI: 00191506043131; Lots: CL12695, CL12696, CL12699, CL12700, CL12703, CL12808, CL12816, CL12817, CL12823, CL12834, CL12835, CL12836, CL12853, CL12860, CL12882, CL12883, CL12884, CL12891, CL12907, CL12908, CL12909, CL12919, CL12949, CL12950, CL12969, CL12976, CL12979, CL12980, CL13005, CL13010, CL13011, CL13012, CL13018, CL13032, CL13033, CL13034, CL13037, CL13056, CL13057, CL13058, CL13074, CL13075, CL13076, CL13077, CL13097, CL13098, CL13099, CL13100, CL13104, CL13115, CL13116, CL13117, CL13118, CL13119, CL13135, CL13136, CL13137, CL13138, CL13139, CL13156, CL13157, CL13191, CL13192, CL13193, CL13194, CL13197, CL13208, CL13209, CL13210, CL13215, CL13216, CL13217, CL13224, CL13229, CL13230, CL13257, CL13258, CL13288, CL13289, CL13337, CL13338, CL13365, CL13366, CL13384, CL13398, CL13399, CL13419, CL13420, CL13423, CL13467, CL13469, CL13470, CL13471, CL13472, CL13517, CL13525, CL13526, CL13527, CL13528, CL13535, CL13536, CL13541, CL13550, CL13551, CL13552, CL13553, CL13555, CL13564, CL13565, CL13566, CL13571, CL13574, CL13575, CL13576, CL13598, CL13599, CL13616, CL13637, CL13642, CL13643, CL13644, CL13665, CL13666, CL13667, CL13668, CL13704, CL13705, CL13706, CL13715, CL13716, CL13721, CL13734, CL13735, CL13736, CL13761, CL13762, CL13763, CL13798, CL13799, CL13800, CL13822, CL13825, CL13826, CL13835, CL13836, CL13837, CL13858, CL13859, CL13860, CL13912, CL13913, CL13914, CL13927, CL13928, CL13929, CL13942, CL13944, CL13945, CL13974, CL13975, CL13981, CL13990, CL13991, CL13992, CL14027, CL14028, CL14029, CL14040, CL14041, CL14042, CL14053, CL14054, CL14055, CL14079, CL14080, CL14081, CL14122, CL14123, CL14124, CL14139, CL14140, CL14141, CL14169, CL14172, CL14173, CL14174, CL14188, CL14189, CL14190, CL14227, CL14228, CL14229, CL14250, CL14251, CL14252, CL14288, CL14289, CL14290, CL14347, CL14348, CL14377, CL14378, CL14398, CL14399, CL14400, CL14440, CL14441, CL14442, CL14489, CL14490, CL14492, CL14493, CL14496, CL14502, CL14505, CL14506, CL14520, CL14583, CL14584, CL14585, CL14634, CL14636, CL14637, CL14671, CL14672, CL14673, CL14674, CL14698, CL14699, CL14700, CL14712, CL14713, CL14714, CL14715, CL14751, CL14752, CL14753, CL14771, CL14772, CL14773, CL14774, CL14815, CL14816, CL14817, CL14818, CL14856, CL14857, CL14858, CL14859, CL14872, CL14873, CL14874, CL14930, CL14931, CL14932, CL14936, CL14937, CL14938, CL14970, CL14971, CL14972, CL15006, CL15007, CL15010, CL15011, CL15068, CL15069, CL15070, CL15080, CL15089, CL15110, CL15130, CL15131, CL15132, CL15194, CL15195, CL15196, CL15241, CL15242, CL15243, CL15273, CL15274, CL15275, CL15276, CL15319, CL15320, CL15321, CL15322, CL15359, CL15360, CL15361, CL15362, CL15398, CL15399, CL15400, CL15401, CL15428, CL15429, CL15430, CL15431, CL15451, CL15475, CL15476, CL15477, CL15491, CL15492, CL15493, CL15494, CL15525, CL15540, CL15541, CL15548, CL15549, CL15570, CL15583, CL15584, CL15599, CL15603, CL15617, CL15641, CL15642, CL15718, CL15756, CL15769, CL15770, CL15777

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 508,151 units.