Recall
FDAHealthMay 13, 2026

Thoratec LLC — Brand Name: HeartMate Product Name: HeartMate 3" System Controller Model/Catalog Number: 106531US Software Version: N/A Product Description: The HeartMate 3" System Controller acts

⚠️ Due to 11 volt Backup Battery failures.

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
May 13, 2026

Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI and the countries of Canada, Germany, Japan, Netherlands, Poland, Spain.

What's affected?

Product
Brand Name: HeartMate Product Name: HeartMate 3" System Controller Model/Catalog Number: 106531US Software Version: N/A Product Description: The HeartMate 3" System Controller acts as the central power and communication hub for the HeartMate 3 LVAS. It passes power from the Power Module, the Mobile Power Unit, Lithium-Ion Batteries, or its own integrated emergency backup supply, down to the LVAD via the Driveline. The HeartMate 3 System Controller constantly monitors system performance through communication with the implanted LVAD and Controller internal measurements and alerts the user to any alarm conditions by activating membrane panel LEDs and integrated audio annunciators. Further information on alarm conditions as well as system status can be attained by the user from the front panel LCD on the System Controller. When connected to a HeartMate Touch" Communication System, the System Controller sends information regarding the System Controller and Pump Status once per second to provide additional information to the user. This link also allows the clinician to set new patient operating parameters (e.g. pump speed) and provides a link for downloading trend and/or event recorder data. Component: Yes, HeartMate 3 Left Ventricular System (LVAS) Kits
Lot code
Model

Additional details

Lot Code: Model Number:106531US UDI/GTIN: 05415067039446 Serial Numbers Distributed in US: See Attachment E for serial numbers distributed.

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 694.

View original FDA notice →

Get Health recall alerts

Delivered to your inbox. Free. No account required.