Recall
FDAHealthJuly 10, 2025

mo-Vis BVBA — Micro Joystick R-net. Electrical wheelchair component.

⚠️ Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
July 10, 2025

US Nationwide distribution in the state of TX.

What's affected?

Product
Micro Joystick R-net. Electrical wheelchair component.
Lot code
Model No. P002-52; UDI-DI: 05407008320027; Serial No. 1000 to 1690.

Additional details

Model No. P002-52; UDI-DI: 05407008320027; Serial No. 1000 to 1690.

What should I do?

Recommended action

Class I recall. Voluntary: Firm initiated Quantity affected: 3 units.