FDAHealthJuly 10, 2025
mo-Vis BVBA — Micro Joystick R-net. Electrical wheelchair component.
⚠️ Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
Is this relevant to me?
- Geography
- Nationwide
- Source agency
- FDA
- Category
- Health
- Published
- July 10, 2025
US Nationwide distribution in the state of TX.
What's affected?
- Product
- Micro Joystick R-net. Electrical wheelchair component.
- Brand
- mo-Vis BVBA
- Lot code
- Model No. P002-52; UDI-DI: 05407008320027; Serial No. 1000 to 1690.
Additional details
Model No. P002-52; UDI-DI: 05407008320027; Serial No. 1000 to 1690.
What should I do?
Recommended action
Class I recall. Voluntary: Firm initiated Quantity affected: 3 units.