VANTIVE US HEALTHCARE LLC — PrisMax V3 control Unit-US, Hemodialysis Delivery System, Product Code 955724

⚠️ Unexpected PrisMax System alarm T0830: Blood Leak Detected, when no blood leak is present, or difficulty normalizing the Blood Leak Detector (BLD) leading to PrisMax System alarms T1313: BLD Normalize Failed, T0853: Normalization Failed, or T1205: BLD Self-Test Failure

Is this relevant to me?

Geography
Unknown
Source agency
FDA
Category
Health
Published
July 28, 2025

Worldwide

What's affected?

Product
PrisMax V3 control Unit-US, Hemodialysis Delivery System, Product Code 955724
Lot code
UDI 07332414126018, All serial numbers

Additional details

UDI 07332414126018, All serial numbers

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 2449 units.