Boston Scientific Corporation — WATCHMAN TruSeal Access System SGL, US, Material Number (UPN) M635TU70010; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Cl
⚠️ Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
Is this relevant to me?
- Geography
- Unknown
- Source agency
- FDA
- Category
- Health
- Published
- July 29, 2025
Worldwide.
What's affected?
- Product
- WATCHMAN TruSeal Access System SGL, US, Material Number (UPN) M635TU70010; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.
- Lot code
- GTIN 08714729965701, ALL NON-EXPIRED BATCHES
Additional details
GTIN 08714729965701, ALL NON-EXPIRED BATCHES
What should I do?
Recommended action
Class I recall. Voluntary: Firm initiated Quantity affected: 340185 units in total.