THOR Photomedicine Ltd — Brand Name: THOR Product Name: NovoTHOR XL Model/Catalog Number: S2190 Software Version: N/A Product Description: The NovoTHOR and NovoTHOR XL are intended to emit energy
⚠️ The potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.
Is this relevant to me?
- Geography
- Nationwide
- Source agency
- FDA
- Category
- Health
- Published
- July 24, 2025
Worldwide distribution: US (nationwide): AK, AZ, CA, CO, CT, FL, GA, HI, ID, IN, MA, MD, NC, ND, NJ, NV, NY, OR, PA, TX, UT, VA, WA, WI, WV; and OUS (International): Australia, United Kingdom, Austria, Ireland, Italy, Lithuania, Netherlands, Poland, Serbia, Slovakia, Spain.
What's affected?
- Product
- Brand Name: THOR Product Name: NovoTHOR XL Model/Catalog Number: S2190 Software Version: N/A Product Description: The NovoTHOR and NovoTHOR XL are intended to emit energy in the visible and IR spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary increase in local blood circulation, temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, muscle spasm, and temporary relaxation of muscles. Component: No
- Lot code
- Model/Catalog Number: S2190; UDI-DI: 05060494130510; Serial Numbers: 4006, 4007, 4008, 4011, 4018, 4023, 4025, 4027, 4031, 4032, 4034, 4039, 4041, 4051, 4052, 4053, 4056, 4057, 4061, 4069, 4075, 4092, 4093, 4098, 4102, 4115, 4119, 4129, 4131, 4135, 4145, 4146, 4147, 4149, 4150, 4154, 4155, 4156, 4160, 4161, 4162, 4014, 4040, 4045, 4060, 4063, 4064, 4067, 4089, 4104, 4108, 4118, 4128, 4136, 4142;
Additional details
Model/Catalog Number: S2190; UDI-DI: 05060494130510; Serial Numbers: 4006, 4007, 4008, 4011, 4018, 4023, 4025, 4027, 4031, 4032, 4034, 4039, 4041, 4051, 4052, 4053, 4056, 4057, 4061, 4069, 4075, 4092, 4093, 4098, 4102, 4115, 4119, 4129, 4131, 4135, 4145, 4146, 4147, 4149, 4150, 4154, 4155, 4156, 4160, 4161, 4162, 4014, 4040, 4045, 4060, 4063, 4064, 4067, 4089, 4104, 4108, 4118, 4128, 4136, 4142;
What should I do?
Recommended action
Class II recall. Voluntary: Firm initiated Quantity affected: 55 units (41 US, 14 OUS).
Get Health recall alerts
Delivered to your inbox. Free. No account required.