Dynex Technologies, Inc. — Dynex Agility Sample Tips; Model Number: 67910; The 67910 Agility Sample Tips are an accessory/consumable used with the Agility Automated ELISA system.

⚠️ The internal label affixed to the sample tip rack is incorrectly labeled as reagent tip racks. When scanned by the Agility instrument, the tip rack is misidentified as a different tip type, rendering it incompatible with the system and preventing use which causes a delay in obtaining patient results.

Is this relevant to me?

Geography
CA, FL, NC, NJ, NY
Source agency
FDA
Category
Health
Published
July 16, 2025

Domestic: CA, FL, NC, NJ, NY;

What's affected?

Product
Dynex Agility Sample Tips; Model Number: 67910; The 67910 Agility Sample Tips are an accessory/consumable used with the Agility Automated ELISA system.
Lot code
863366

Additional details

Model Number: 67910; UDI-DI: 5060456180072; Lot Number: 863366;

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 237 units.