FDAHealthJuly 2, 2025
GE Medical Systems, LLC — GE HealthCare OEC 9800.
⚠️ GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.
Is this relevant to me?
- Geography
- Nationwide
- Source agency
- FDA
- Category
- Health
- Published
- July 2, 2025
US Nationwide distribution.
What's affected?
- Product
- GE HealthCare OEC 9800.
- Lot code
- GTIN: 00840682114349
Additional details
GTIN: 00840682114349
What should I do?
Recommended action
Class II recall. FDA Mandated
Get Health recall alerts
Delivered to your inbox. Free. No account required.