Volcano Corp — Volcano Visions Digital IVUS Catheter: PV.014P (Platinum), REF:85910P; PV.014P RX, REF: 014R; and PV.018, REF: 86700, 86700J used with Instructions for use.

⚠️ Reports of IVUS catheter and guide wire entanglement during radial-to-peripheral procedures, due to catheter use without appropriate sheath and/or guide catheter, which may require further intervention, such as surgical removal; Firm is restating instructions for use and adding: "Use a guide sheath of appropriate length to provide adequate support to the rapid exchange IVUS catheter and guidewire"

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
June 13, 2025

Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of BR, PL, SK, IT, DK, DE, GB, AT, KW, CR, ES, NL, LV, MT, IE, CH, ZA, RO, FI, PR, FR, CA, PA, IL, SE, PT, HR, GR, NZ, HK, TW, MY, TH, MX, AR, PE, AE, BG, AU, JP, ID, RS, CO, SG, BE.

What's affected?

Product
Volcano Visions Digital IVUS Catheter: PV.014P (Platinum), REF:85910P; PV.014P RX, REF: 014R; and PV.018, REF: 86700, 86700J used with Instructions for use.
Lot code
UDI-DI: 00845225002848, 00184360000037. Instructions for use: 300004875732_A Revision Date: 03/2021, 300004410212.A Release Date: 03/2021

Additional details

UDI-DI: 00845225002848, 00184360000037. Instructions for use: 300004875732_A Revision Date: 03/2021, 300004410212.A Release Date: 03/2021

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 233,817.