Recall
FDAHealthMarch 9, 2026

Siemens Medical Solutions USA, Inc — ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model 11327600. ARTIS icono ceiling Model 11328100. ARTIS icono floor Model 11327700. interventional f

⚠️ During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

Is this relevant to me?

Geography
US
Source agency
FDA
Category
Health
Published
March 9, 2026

US and Worldwide

What's affected?

Product
ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model 11327600. ARTIS icono ceiling Model 11328100. ARTIS icono floor Model 11327700. interventional fluoroscopic x-ray system
Lot code
ARTIS icono biplane UDI 4056869063317 ARTIS icono ceiling UDI 4056869295923 ARTIS icono floor UDI 4056869149325

Additional details

ARTIS icono biplane UDI 4056869063317 ARTIS icono ceiling UDI 4056869295923 ARTIS icono floor UDI 4056869149325

What should I do?

Recommended action

Class II recall. FDA Mandated

View original FDA notice →

Get Health recall alerts

Delivered to your inbox. Free. No account required.