Recall
FDAHealthMay 30, 2025

Remote Diagnostic Technologies Ltd. — Inseego USB8 4G Dongle Kit, Part Number: 01-2298, which contains USB Flash Drive, Part Number: 43-2011. The kit is an accessory to the Tempus Pro Monitor, REF: 00

⚠️ Kit USB flash drive contains outdated software, which could result in reverting patient monitor to outdated software and reintroducing the video laryngoscope issue (related to an August 2023 recall), which could lead to delay in diagnosis, delayed treatment, hypoxia due to unexpected loss of video laryngoscopy and all other monitor measurements during system restart.

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
May 30, 2025

Worldwide - US Nationwide distribution in the states of SC, KY, NV, IA, MI and the country of GB.

What's affected?

Product
Inseego USB8 4G Dongle Kit, Part Number: 01-2298, which contains USB Flash Drive, Part Number: 43-2011. The kit is an accessory to the Tempus Pro Monitor, REF: 00-1007-R, 00-1024-R, 00-1026-R
Lot code
REF/UDI-DI: 00-1007-R/05060472440013, 05060472442918; 00-1024-R/05060472441027, 05060472442925; 00-1026-R/05060472441058, 05060472442932. Software Version: vx.28

Additional details

REF/UDI-DI: 00-1007-R/05060472440013, 05060472442918; 00-1024-R/05060472441027, 05060472442925; 00-1026-R/05060472441058, 05060472442932. Software Version: vx.28

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 55.

View original FDA notice →

Get Health recall alerts

Delivered to your inbox. Free. No account required.