Recall
FDAHealthJune 12, 2025

Beckman Coulter Inc. — Estrone RIA, REF: DSL8700. Estrone RIA is an IVD used for the quantitative measurement of estrone in human serum and plasma.

⚠️ A specific lot of bovine serum albumin (BSA) used for manufacturing of coated tubes (CT) that are included in the Estrone RIA kit lots caused different affinity of manufactured coated tubes to patient samples, causing falsely increased assay results for the impacted Estrone RIA kits.

Is this relevant to me?

Geography
US
Source agency
FDA
Category
Health
Published
June 12, 2025

US: North Carolina OUS: Argentina, Austria, Belgium, Canada, France, Germany, Greece, Hungary, Italy, Netherlands

What's affected?

Product
Estrone RIA, REF: DSL8700. Estrone RIA is an IVD used for the quantitative measurement of estrone in human serum and plasma.
Lot code
250106D

Additional details

UDI: 15099590211615/Lot: 250106D, 250203C, 250203D, 250303C, 250331C, 250428C, 250526C

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 496 units.

View original FDA notice →

Get Health recall alerts

Delivered to your inbox. Free. No account required.