Civco Medical Instruments Co. Inc. — eTRAX Needle System Starter Kit 16G(for Aurora Trackers), Part Number 667-151
⚠️ There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
Is this relevant to me?
- Geography
- US, OH, PA, WA
- Source agency
- FDA
- Category
- Health
- Published
- March 2, 2026
US States: OH, PA, WA. China.
What's affected?
- Product
- eTRAX Needle System Starter Kit 16G(for Aurora Trackers), Part Number 667-151
- Lot code
- A228675
Additional details
UDI-DI: 00841436111058; Lot Number: A228675, A228676, A233854, A268115, A268237, A272509, A277538, A280341
What should I do?
Recommended action
Class II recall. Voluntary: Firm initiated Quantity affected: 18 units.