I.T.S. GmbH — Pelvic Reconstruction System (PRS) with the below descriptions and corresponding article numbers. 1. Fixation Screw; Article Number: 70312. 2. Spike short; Article Number: 70314.

⚠️ Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
January 23, 2026

US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

What's affected?

Product
Pelvic Reconstruction System (PRS) with the below descriptions and corresponding article numbers. 1. Fixation Screw; Article Number: 70312. 2. Spike short; Article Number: 70314.
Lot code
):

Additional details

1. Article Number (Lot Number): 70312 (26/F65133). 2. Article Number (Lot Number): 70314 (86/619736).

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 11 units.