I.T.S. GmbH — Brand Name: Humeral Head Plate with Angular Stability, with the following description: Proximal Humeral Plate. Article Numbers: 211341-10, 211341-4, 211341-7, 21134-4, 21134-5, 21135-4,

⚠️ Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
January 23, 2026

US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

What's affected?

Product
Brand Name: Humeral Head Plate with Angular Stability, with the following description: Proximal Humeral Plate. Article Numbers: 211341-10, 211341-4, 211341-7, 21134-4, 21134-5, 21135-4, 21136-4, 21136-5, 21136-7, 21137-4, 21137-7, 21138-4, 21139-4.
Lot code
):

Additional details

Article Number (Lot Numbers): 211341-10 (26/F63368, 26/F65625, 26/F66904, 26/F669041, 679/012410), 211341-4 (26/F63366, 26/F65626, 26/F656261, 26/F66145, 26/F661451, 26/F66496, 26/F66905, 26/F669051, 26/F73497, 679/012439, 679/022450), 211341-7 (26/F63367, 26/F65627, 26/F66146, 26/F66906, 679/012440), 21134-4 (768/18758, 768/19642), 21134-5 (778/201904), 21135-4 (26/F67801, 768/21382), 21136-4 (26/F65629, 26/F66907, 26/F67802, 26/F70071, 768/21383), 21136-5 (770/101901), 21136-7 (26/F65039, 26/F65630), 21137-4 (26/F65631, 26/F67804, 768/21384), 21137-7 (26/F65040), 21138-4 (26/F65633, 26/F66908, 26/F67806), 21139-4 (26/F65634, 26/F67807, 768/21386).

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 292 units.