Olympus Corporation of the Americas — Brand Name: Olympus ShockPulse-SE Lithotripsy System with Generator Product Name: ShockPulse-SE Lithotripsy System - Reuseable Probes Model/Catalog Number: SPL-

⚠️ Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment. It was discovered that additional serial numbers of ShockPulse generators may also have the mis-wired component. As a result, this is expanding the action to the additional impacted devices.

Is this relevant to me?

Geography
Unknown
Source agency
FDA
Category
Health
Published
January 8, 2026

International distribution in the country of Canada, Germany, Singapore, Australia, and India.

What's affected?

Product
Brand Name: Olympus ShockPulse-SE Lithotripsy System with Generator Product Name: ShockPulse-SE Lithotripsy System - Reuseable Probes Model/Catalog Number: SPL-SR, contains generator SPL-G Product Description: An assembly of devices that uses a combination of, or individually applied, pneumatically-created ballistic shock waves and ultrasonic waves for the intracorporeal fragmentation of stones (calculi) in the urinary tract (i.e., kidney, ureter, and bladder) for their removal. It typically includes a pressure and power regulator/control unit, handpieces, and two types of probes of various sizes. The probes are attached to the two handpieces which are connected to the control unit that supplies the energy for both the pneumatic and ultrasonic functions. Energy is applied to a calculus which is fragmented in situ and removed via a suction system facilitated by the lithotripsy system.
Lot code
Model/Catalog Number: SPL-SR, contains generator SPL-G; UDI: SPL-SR 00821925043824, SPL-G 00821925044203; Serial Numbers: CG5026, CG5027, CG5028, CG5029, CG5030, CG5031, CG5032, CG5033, CG5034, CG5035, CG5036, CG5037, CG5039, CG5040, CG5041, CG5042, CG5043, CG5045, CG5046, CG5047, CG5048, CG5049, CG5050, CG5051, CG5052, CG5053, CG5054, CG5055, CG5056, CG5057, CG5058, CG5059, CG5060, CG5061, CG5062, CG5063, CG5065, CG5066, CG5067, CG6037, CG6061, CG6069, CG6070, CG6071, CG6088, CG6094, CG6095, CG6096, CG6097, CG6098, CG6099, CG7000, CG7001, CG7002, CG7003;

Additional details

Model/Catalog Number: SPL-SR, contains generator SPL-G; UDI: SPL-SR 00821925043824, SPL-G 00821925044203; Serial Numbers: CG5026, CG5027, CG5028, CG5029, CG5030, CG5031, CG5032, CG5033, CG5034, CG5035, CG5036, CG5037, CG5039, CG5040, CG5041, CG5042, CG5043, CG5045, CG5046, CG5047, CG5048, CG5049, CG5050, CG5051, CG5052, CG5053, CG5054, CG5055, CG5056, CG5057, CG5058, CG5059, CG5060, CG5061, CG5062, CG5063, CG5065, CG5066, CG5067, CG6037, CG6061, CG6069, CG6070, CG6071, CG6088, CG6094, CG6095, CG6096, CG6097, CG6098, CG6099, CG7000, CG7001, CG7002, CG7003;

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 55 units (All OUS).