Recall
FDAHealthJanuary 21, 2026

Agiliti Health - Ellis — Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt

⚠️ Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function.

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
January 21, 2026

US Nationwide distribution.

What's affected?

Product
Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure
Lot code
UDI-DI 0084569904914; All serial numbers with a manufacturing date prior to 12/04/2025

Additional details

UDI-DI 0084569904914; All serial numbers with a manufacturing date prior to 12/04/2025

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 4286.

View original FDA notice →

Get Health recall alerts

Delivered to your inbox. Free. No account required.