Recall
FDAHealthJanuary 13, 2026

Laerdal Medical Corporation — Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; 3) LCSU

⚠️ units manufactured in a limited time period may emit electromagnetic noise beyond the acceptable limits for RTCA application specified in RTCA DO-160G Chapter 21, Equipment Category M.

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
January 13, 2026

Worldwide distribution - US Nationwide and the countries of Denmark, Sweden, Norway, Canada, Finland, Australia, Spain, France, United Kingdom, Italy, Korea, Germany, Netherlands, Switzerland, New Zealand, Singapore, Malaysia, Austria.

What's affected?

Product
Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; 3) LCSU 4, RTCA; Catalog Number: 881152;
Lot code
1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; UDI-DI: 07045432088519; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; UDI-DI: 07045432088533; 3) LCSU 4, RTCA; Catalog Number: 881152; UDI-DI: 07045432067132;

Additional details

1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; UDI-DI: 07045432088519; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; UDI-DI: 07045432088533; 3) LCSU 4, RTCA; Catalog Number: 881152; UDI-DI: 07045432067132;

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 1202 units (537 US, 665 OUS) (2/23/2026: 574 US, 1239 units).