Recall
FDAHealthDecember 11, 2025

Fujirebio Diagnostics, Inc. — Product Name: Lumipulse G pTau 217 Plasma Immunoreaction Cartridges Model/Catalog Number: 81550 Software Version: Not Applicable Product Description: Lumipulse G pTau

⚠️ Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
December 11, 2025

US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.

What's affected?

Product
Product Name: Lumipulse G pTau 217 Plasma Immunoreaction Cartridges Model/Catalog Number: 81550 Software Version: Not Applicable Product Description: Lumipulse G pTau 217 Plasma Immunoreaction Cartridges (IRC): Lumipulse G pTau 217 Plasma is an assay system, including a set of immunoassay reagents (Lumipulse G pTau 217 Plasma IRC), for the quantitative measurement of pTau 217 in K2EDTA plasma specimens based on chemiluminescent enzyme immunoassay (CLEIA) technology by a specific two-step immunoassay method on the LUMIPULSE G System. 3 ¿ 14 Tests: 3 racks, 14 IRCs per rack, 1 test per IRC. Each IRC contains (abbreviated contents): 1. Antibody-Coated Particle Solution: (Contains anti-phosphorylated Tau (217) monoclonal antibody (mouse)-coated particles). This solution contains gelatin and turns into gel at 15¿C or lower, 2. Enzyme-Labeled Antibody Solution: (Contains alkaline phosphatase (ALP)-labeled anti-Tau monoclonal antibodies (mouse) conjugate), 3. Assay Specific Solution: (Contains chemical stabilizers in 50 mM Tris buffer) Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio
Lot code
D4C6028U

Additional details

Model/Catalog Number: 81550; UDI-DI: 05414470815500; Lot Numbers: D4C6028U, D4C6029U, D4C6051U, D4C6052U;

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 1569 units.