Recall
FDAHealthNovember 28, 2025

RAYSEARCH LABORATORIES AB — RayStation/RayPlan. Model/Catalog Numbers: 11.0.0, 11.0.1, 11.0.3, 11.0.4. Software Version: RayStation 11A, RayStation 11A SP1, RayStation 11A SP2, RayStation 11A SP3.

⚠️ Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

Is this relevant to me?

Geography
US
Source agency
FDA
Category
Health
Published
November 28, 2025

Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS

What's affected?

Product
RayStation/RayPlan. Model/Catalog Numbers: 11.0.0, 11.0.1, 11.0.3, 11.0.4. Software Version: RayStation 11A, RayStation 11A SP1, RayStation 11A SP2, RayStation 11A SP3. Radiation Therapy Treatment Planning System
Lot code
UDI:

Additional details

Lot Code: UDI: 0735000201038920210518, 0735000201043320210610, 0735000201044020210916, 35000201063120220616. GTIN: 07350002010389, 07350002010433, 07350002010440, 07350002010631. Serial Numbers: 11.0.0.951, 11.0.1.29, 11.0.3.116, 11.0.4.15. Software Revision: RayStation 11A, 11A SP1, 11A SP2, 11A SP3. Expiration Date: 2027-06-22.

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 10 units.

View original FDA notice →

Get Health recall alerts

Delivered to your inbox. Free. No account required.