Recall
FDAHealthDecember 5, 2025

Beaver-Visitec International, Inc. — BVI Leos Laser and Endoscopy System. Model Number: OME6000U.

⚠️ Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
December 5, 2025

US Nationwide distribution in the states of FL, MI, OK, TX, WI.

What's affected?

Product
BVI Leos Laser and Endoscopy System. Model Number: OME6000U.
Lot code
Model Number: OME6000U. UDI: (01)00886158037576(11)250513(21). Serial Numbers: 77UBV500003, 77UBV500004, 77UBV500005, 77UBV500006, 77UBV500010.

Additional details

Model Number: OME6000U. UDI: (01)00886158037576(11)250513(21). Serial Numbers: 77UBV500003, 77UBV500004, 77UBV500005, 77UBV500006, 77UBV500010.

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 5 units.

View original FDA notice →

Get Health recall alerts

Delivered to your inbox. Free. No account required.