DICOM Grid, Inc. — Intelerad InteleShare software, with ProViewer component

⚠️ Software intended to aid in diagnosing conditions, planning treatments, visualizing anatomical structures has a bug that, if all of following are met: Viewing images in InteleShare viewer; Multiplanar reconstruction applied; Manual rotation applied, could result in inaccurate length measurement tool values that could compromise diagnostic accuracy, lead to misdiagnosis or inappropriate follow-up.

Is this relevant to me?

Geography
US, AZ, TX, GA, NV, MO, MI, IL, OH, CT, OR, FL, NY, NM, MN, PA, AK, ND, NJ, VA, WA, NC, TN, WI, CA, SD, CO, MA, WV, KS, MD, VT, IA, RI, AL, ID, LA, PR, NH, IN, OK, UT, NE, SC, AR, DE, NZ, AU, GB, BE, FR, IT, NO, SE
Source agency
FDA
Category
Health
Published
November 18, 2025

US: AZ, TX, GA, NV, MO, MI, IL, OH, CT, OR, FL, NY, NM, MN, PA, AK, ND, NJ, VA, WA, NC, TN, WI, GA, AK, CA, SD, CO, MA, WV, MN, KS, MD, VT, IA, RI, AL, ID, LA, PR, NH, IN, OK, UT, NE, SC, AR, DE OUS: CA, NZ, AU, GB, BE, FR, IT, NO, SE

What's affected?

Product
Intelerad InteleShare software, with ProViewer component
Lot code
Software versions 3.24.3 - 3.25.2. Date of First Distribution: 2024-08-28. UDI-DI: B228INTELESHARE0

Additional details

Software versions 3.24.3 - 3.25.2. Date of First Distribution: 2024-08-28. UDI-DI: B228INTELESHARE0

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 924.