Bard Peripheral Vascular Inc — Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,
⚠️ Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt position, which may lead users to believe device is not yet in pleural or peritoneal space when it is, may cause discomfort, delay to treatment, injury to internal organs or vessels.
Is this relevant to me?
- Geography
- Nationwide
- Source agency
- FDA
- Category
- Health
- Published
- November 6, 2025
US Nationwide distribution in the states of CA, MN, IN, MI, LA.
What's affected?
- Product
- Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,
- Lot code
- /Expiration:
Additional details
UDI/Lot/Expiration: (01)10885403108365(17)271130(10)0001623320/0001623320/20271130; (01)10885403108365(17)271130 (10)0001623069/0001623069/20271130
What should I do?
Recommended action
Class II recall. Voluntary: Firm initiated Quantity affected: 1240.
Get Health recall alerts