Philips North America Llc — CT 5300; Software Version Number: 4.5, 5.0, 5.1;
⚠️ If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems during rotation. Other components could be damaged due to the contact with the tube heat exchanger.
Is this relevant to me?
- Geography
- Nationwide
- Source agency
- FDA
- Category
- Health
- Published
- October 29, 2025
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Bosnia-Herz., Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Ethiopia, France, FrenchPolynesia, Georgia, Germany, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Lao, Latvia, Lebanon, Lesotho, Libya, Lithuania, Macedonia, Malaysia, Mauritius, Mexico, Moldova, Montenegro, Montserrat, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Papua Nw Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Rwanda, Saint Kitts, Saint Lucia, Saudi Arabia, Serbia, Sint Maarten (D, Slovakia, South Africa, South Korea, Spain, St.Pier,Miquel., Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Trinidad,Tobago, Tunisia, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
What's affected?
- Product
- CT 5300; Software Version Number: 4.5, 5.0, 5.1;
- Lot code
- 1) Model Number: 728285; UDI-DI: 00884838113237; Serial Numbers: 35071, 800007, 800002, 800001, 800004, 800003;
Additional details
1) Model Number: 728285; UDI-DI: 00884838113237; Serial Numbers: 35071, 800007, 800002, 800001, 800004, 800003;
What should I do?
Recommended action
Class II recall. Voluntary: Firm initiated Quantity affected: 6 units (OUS only).