Recall
FDAHealthOctober 13, 2025

Paragon 28, Inc. — Phantom Fibula Nail, 2.8mm x 130mm, Right, Ti, Fibula Nail, REF: P36-128-130R-S

⚠️ Fibula nail system intended for use in fixation of fibular fractures and osteotomies, has out of specification thread depths on the fibula nail, so implant may not be able to mate properly with the associated mounting bolt. If nail (implant) cannot properly mate with instrumentation, the surgery potentially will need to be aborted.

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
October 13, 2025

US Nationwide distribution in the states of NJ, WA, NY, MD, CA, FL, NM, TX, IL, OH, NV, TN, LA.

What's affected?

Product
Phantom Fibula Nail, 2.8mm x 130mm, Right, Ti, Fibula Nail, REF: P36-128-130R-S
Lot code
501182207125

Additional details

UDI-DI: 00889795133670. Lot: 501182207125. Expiration: March 13th, 2030

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 20.

View original FDA notice →

Get Health recall alerts

Delivered to your inbox. Free. No account required.